DHG-9030A series electric blast drying oven self-tuning function instructions

Auto-tuning <br> <br> if the temperature control process, a larger temperature overshoot, or large temperature fluctuations, press the following auto tuning function a. Turn off the power switch, open the door, so that The equipment is naturally cooled to ambient temperature;

b. Close the door, open the power switch, set the temperature to the usual temperature value;

c. Under normal working conditions, press and hold the shift button for 5 seconds, then enter the auto-tuning state, the RUN/AT indicator flashes;

d. After the auto-tuning is finished, the meter will automatically return to normal working condition.

e. In the auto-tuning state, pressing any key is invalid.

Note: During the control parameter adjustment process, if no key is pressed for 6 seconds, the meter will automatically return to normal operation, but the changed data is considered invalid.

Since the product has been rigorously tested before leaving the factory, it is generally not necessary to make corrections. However, the product is used for a period of time; the product is used for a period of time; the use environment is bad; the ambient temperature used is inconsistent with the ambient temperature during manufacturing; when the value of the control temperature is changed before and after, etc., the function parameters can be adjusted to meet the factory standard. .

One-Step Method Test Solution Of COVID-19

Uni-medica provides a rapid solution for screening and detecting COVID-19. It takes only 40 minutes for one round, the instruments and reagents used in this solution including sample tube, sample released reagent,PCR test reagent, PCR system.This solution doesn't need Nucleic Acid Extractor, and the results can be output on-site.